CP103459 | Extensometer Digital Calibrator

Model 3590VHR High Resolution Extensometer Calibrator

  • Scale resolution: 20 nm (0.8 microinches);
  • Scale Accuracy:
  • ±0.19 microns (±7.5 microinches) over any 0.12 mm (0.00475 inch) range.
  • Meets ASTM E83 Class B-1 for a 6 mm gauge length and greater extensometers and ISO 9513 Class 0,5 over the full measuring range of the calibrator.
  • 50 mm (2 inches) measuring range;
  • USB port, Netbook provided;
  • Uses the same adapters as the model 3590 and 3590HR calibrators;
  • Provided with a foam lined case for storage;
  • Comes with 9.5 mm diameter round posts for extensometer attachment;
  • Includes ISO 17025 accredited NPL (UK) traceable calibration certificate for the full calibrator stand.
  • 3590-02 Adapters for flat specimens
  • 3590-03 Adapters for clip-on fracture mechanics gauges
  • 3590-04 Transverse extensometer adapters

*Note: ISO527 requires 1 micron accuracy extensometer.

Calibrator resolution should be better by at least a factor of two.

ASTM E399 requires 20 uin resolution

Read More ...