A Guide to Cyclic Fatigue Testing of Artificial Hip Implant Prostheses in accordance with ISO 7206-4, ISO 7206-6, ISO-7206-8 and ASTM F2068
- Biomedical
- BioMaterials
- ASTM F2068
- ISO 7206-4
- ISO 7206-6
- ISO 7206-8
- Compression
- High Cycle Fatigue
- Multi-Axis
- Shear
- Torsion
Written by Toby Lane
A SUMMARY OF THE STANDARDS AND THE DYNAMIC TESTS INVOLVEDImplanted metal devices can often experience loosening within the host bone (proximal loosening) as a result of stress shielding. Stress shielding can be described as localized degradation in bone strength due to a decrease in physiological loading of certain areas because of the presence of the stiffer metal implant. As this can occur following even normal activity, Fatigue testing of hip implants is required to understand how abnormal loading profiles can arise and to evaluate endurance properties by simulating the dynamic loading of the implant during gait.The ISO and ASTM standards have been established to test for both abnormal and normal fatigue loading.
- ISO 7206-4: Simulates loading when proximal loosening has occurred. Loads are applied through the femoral head of the hip implant to induce compressive, bending and torsional stresses.
- ISO 7206-6: Examines fatigue of the implant neck, which is more consistent with a correctly fixed implant subjected to normal in vivo loading.
- ISO 7206-8: Specifies the endurance performance of the implant with the application of torsion.
- ASTM F2068: “Standard Specification for Femoral Prostheses—Metallic Implants” describes hip implant specifications with reference to the ISO standards.
- Achieving precise orientation of the specimen in the embedding material due to the specimen geometry.
- Avoiding heating of specimen due to high frequency testing.
- Set-up must allow for acceptable compressive, flexural and torsional strain in the specimen.


- A lower specimen holder or housing that will contain an embedding medium. The medium used should have a modulus of elasticity between 2000 and 6000 N/mm2 (typically acrylic bone cement or epoxy resin are used).
- A fluid container with temperature control capability, large enough to fully submerge the specimen. Typically, a 0.9 g/l saline solution is used (NaCl + deionized water).
- A device for gripping the head or neck of the specimen, maintaining the specimen orientation.
- Adapter for transferring the machine load to the imbedded specimen. The device must feature a lower friction mechanism that is designed to reduce loads that aren’t coincident with the loading axis of the test machine.

An R ratio of 0.1 is typically used for hip implant testing. As an example, ASTM F2068 suggests testing components between 534N and 5340N.
Dispositivo de fatiga para articulación de cadera, ISO 7206
El dispositivo de fatiga para articulación de cadera se ha diseñado específicamente para cumplir con los requisitos mejorados de la norma ISO 7206-4. El dispositivo simula la carga de fatiga de un vástago de cadera durante un ciclo de marcha.
Catálogo de ElectroPuls®
Los instrumentos ElectroPuls son sistemas avanzados de ensayo electrodinámico para la realización de pruebas dinámicas y estáticas de materiales y componentes. Alimentados totalmente por una red monofásica, cuentan con la última tecnología para ensayos y otras muchas funciones orientadas al usuario.
Catálogo WaveMatrix™2
WaveMatrix2 es un software inteligente diseñado para los ensayos de fatiga y dinámicos de materiales y componentes. Ofrece flexibilidad: realiza todo tipo de ensayos, desde una simple rampa estática hasta formas de onda cíclicas y complejas pruebas multiaxiales. El entorno altamente visual, con pantallas tabulares integradas, estructuras de menú claras, vista previa de ensayos matriciales basados en el tiempo y un espacio de trabajo de ensayos en vivo configurable, está diseñado para ser intuitivo y fiable. Ofrece varias funciones inteligentes, como la reducción de datos y la organización de proyectos integrada, diseñadas para simplificar tus ensayos.