ASTM F2458 Wound Closure Strength of Tissue Adhesives and Sealants

Mechanical Testing of Bioadhesives

ASTM F2458

ASTM F2458 is a testing standard that specifically focuses on characterizing the wound closure strength of tissue adhesives or sealants. Adhesive wound closure is sometimes preferable to using staples or stitches, as it reduces patient discomfort is and is often more easy to apply. It is extremely important that these products perform as expected, as failure could cause internal or external bleeding and leave a patient vulnerable to infection at the wound site. In order for wounds to heal properly over time and prevent scarring, the adhesive must continue to hold the tissue together until healing occurs.

ASTM F2458 is primarily intended for comparative testing between different adhesive formulations. Identifying a repeatable and life-like substrate enables this standard to be used for quality control purposes of adhesive wound closure products. The test will provide the maximum force at the time of failure as well as the failure mode – typically labelled as adhesive (failure of the adhesive/substrate bond), cohesive (failure of the internal adhesive bond), or substrate failure (failure of the tissue). To read the standard in full, purchase ASTM F2458.

Materials Testing System

For ASTM F2458 testing we recommend a 6800 Series single column testing system with a 100 N load cell and pneumatic grips. Because tissues are typically fragile, they require surfalloy style faces for clamping on the specimen successfully. A flexible coupling is required to ensure the loading is applied through the center of the specimen. For closure devices used internally within the body, the testing should be performed at a temperature of 37 Celsius. A biobox is useful for testing at body temperature.

ASTM F2458 Test Setup
1) 5900 Series Universal Testing System
2) 2530-100 Load Cell
3) Pneumatic Side Action Grips
4) Instron BioBox

 

ASTM F2458 system setup
Related Content

Serie 6800 - Opuscolo dei sistemi di test Premier

I sistemi di test universali della serie 6800 di Instron offrono precisione e affidabilità impareggiabili. Grazie alla serie 6800, costruita sulla base di un’architettura di sistema in attesa di brevetto e dotata delle nuovissime funzionalità Smart-Close Air Kit e Collision Mitigation, i test dei materiali non sono mai stati così semplici, intelligenti e sicuri.

Opuscolo su Bluehill Universal

Bluehill Universal Software è realizzato appositamente per l’interazione touch e un’esperienza utente intuitiva. Scopri test più semplici e più intelligenti con funzionalità quali i metodi di test precaricati, QuickTest in pochi secondi, esportazione di dati migliorata e Instron Connect, una nuova caratteristica che fornisce un collegamento diretto di comunicazione con l’Assistenza. Gli utenti delle versioni precedenti del software come Bluehill 2 e Bluehill 3 possono facilmente aggiornarle alla versione più recente di Bluehill.

BioBox per test di dispositivi medici a 37  C

BioBox di Instron fornisce un ambiente ad aria controllata a temperatura corporea (37 °C) per test pratici di dispositivi medici e biomateriali.

Prese pneumatiche a immersione ad azione laterale e BioBath a temperatura controllata

Il BioBath di Instron a temperatura controllata permette una simulazione accurata delle condizioni ambientali necessarie per il test biomedico.