A SUMMARY OF THE STANDARDS AND THE DYNAMIC TESTS INVOLVEDImplanted metal devices can often experience loosening within the host bone (proximal loosening) as a result of stress shielding. Stress shielding can be described as localized degradation in bone strength due to a decrease in physiological loading of certain areas because of the presence of the stiffer metal implant. As this can occur following even normal activity, Fatigue testing of hip implants is required to understand how abnormal loading profiles can arise and to evaluate endurance properties by simulating the dynamic loading of the implant during gait.The ISO and ASTM standards have been established to test for both abnormal and normal fatigue loading.
- ISO 7206-4: Simulates loading when proximal loosening has occurred. Loads are applied through the femoral head of the hip implant to induce compressive, bending and torsional stresses.
- ISO 7206-6: Examines fatigue of the implant neck, which is more consistent with a correctly fixed implant subjected to normal in vivo loading.
- ISO 7206-8: Specifies the endurance performance of the implant with the application of torsion.
- ASTM F2068: “Standard Specification for Femoral Prostheses—Metallic Implants” describes hip implant specifications with reference to the ISO standards.
- Achieving precise orientation of the specimen in the embedding material due to the specimen geometry.
- Avoiding heating of specimen due to high frequency testing.
- Set-up must allow for acceptable compressive, flexural and torsional strain in the specimen.


- A lower specimen holder or housing that will contain an embedding medium. The medium used should have a modulus of elasticity between 2000 and 6000 N/mm2 (typically acrylic bone cement or epoxy resin are used).
- A fluid container with temperature control capability, large enough to fully submerge the specimen. Typically, a 0.9 g/l saline solution is used (NaCl + deionized water).
- A device for gripping the head or neck of the specimen, maintaining the specimen orientation.
- Adapter for transferring the machine load to the imbedded specimen. The device must feature a lower friction mechanism that is designed to reduce loads that aren’t coincident with the loading axis of the test machine.

An R ratio of 0.1 is typically used for hip implant testing. As an example, ASTM F2068 suggests testing components between 534N and 5340N.
Dispositivo per affaticamento femorale per anca ISO 7206
Il dispositivo per affaticamento femorale per anca è stato progettato specificamente per soddisfare i requisiti avanzati della norma ISO 7206-4. Il dispositivo simula il carico di fatica di uno stelo dell’anca durante un ciclo di deambulazione.
Opuscolo su ElectroPuls®
Gli strumenti ElectroPuls sono sistemi di test elettrodinamici avanzati utilizzabili per test dinamici e statici su materiali e componenti. Alimentati interamente da rete monofase, sono dotati delle più recenti tecnologie per test e di una serie di altre funzionalità concepite per l’utente.
Opuscolo su WaveMatrix™2
WaveMatrix2 è un software intelligente progettato per prove di fatica e dinamiche di materiali e componenti. È estremamente flessibile e consente di testare di tutto: da una semplice rampa statica alle forme d’onda cicliche, fino all’esecuzione di complessi test multifase e multiassiali. L’ambiente visivo molto chiaro, con schermate tabulari integrate, menu chiaramente strutturati, anteprime di test a matrice basati sul tempo e aree di lavoro per test in tempo reale configurabili è stato progettato per essere intuitivo e infondere fiducia. Dotato di funzionalità intelligenti, come la riduzione dei dati e l’organizzazione di progetti integrata, è concepito per rendere più semplice la conduzione di test.