BS EN 868-5 peel test of medical packaging

plastics hub icon 

EN 868-5 characterizes the seal strength of single-use sterilized medical packaging. This standard specifically focuses on packaging materials and systems for medical devices which need to be sterilized. The seal strength identifies the adhesion strength between the two materials of the packaging; typically, a coated paper backing, and a plastic film overlay. This standard is technically equivalent to ASTM F88 with minor differences in the specimen preparation and test procedure.

While packaging can span many industries, The seal strength of medical packaging is extremely important for two reasons – first to ensure that the packing can be opened easily by the medical professional, and second to ensure that the adhesive is strong enough to prevent any breaches of the packaging prior to their use. During shipping and storage the adhesive must remain viable to maintain sterilization: any breach in the packaging would desterilize the product potentially leading to life threatening infections for the patient.

EN 868-5 Test Setup EN 868-5 can be performed on an Instron 3400 or 6800 Series single column universal testing system with a 500 N load cell.



EN 868-5 Test Setup
3400 Series Universal Testing System
Bluehill Universal Software
2530 Series Load Cell
2710-112 Screw Side Action Grips

 

EN 868-5 peel test of sterile medical packaging
Grips

When testing an adhesive to a thick substrate, standard grips result in a misaligned specimen. Testing to EN 868-5 requires either advanced screw side action grips or pneumatic side action grips, both of which can be adjusted to offset to ensure that the specimen remains centered in the load string. Pneumatic grips provide an advanced solution by ensuring a repeatable gripping force to eliminate slippage specimen slippage. Both options can be fitted with specimen centering devices to ensure proper alignment for each test, and both employ a patented “quick-change” jaw face design which enables operators to easily change jaw faces to accommodate the needs of their materials. Rubber-coated grip faces should be used so as to avoid tearing the delicate specimen. Jaw faces are available in many different sizes and are easily interchangeable. 

250 N Model | 2712-052
2710-112
500 N Model | 2710-112
Tips and Tricks

The specimen width and test speed are different than ASTM F88, but produce statistically similar results when the data is normalized.

Related Content

Opuscolo su Bluehill Universal

Bluehill Universal Software è realizzato appositamente per l’interazione touch e un’esperienza utente intuitiva. Scopri test più semplici e più intelligenti con funzionalità quali i metodi di test precaricati, QuickTest in pochi secondi, esportazione di dati migliorata e Instron Connect, una nuova caratteristica che fornisce un collegamento diretto di comunicazione con l’Assistenza. Gli utenti delle versioni precedenti del software come Bluehill 2 e Bluehill 3 possono facilmente aggiornarle alla versione più recente di Bluehill.

Prese pneumatiche ad azione laterale serie 2712-04x

Le prese pneumatiche ad azione laterale forniscono una soluzione versatile per la presa di una vasta gamma di materiali. Capacità nominali tra 1 e 10 kN.

Serie 2710 - Prese a vite ad azione laterale

Le prese a vite ad azione laterale forniscono un metodo molto semplice ed efficiente per tenere i provini di test e possono essere utilizzate in una vasta gamma di applicazioni.