ISO 7886-1 살균된 피하주사기의 시험

 ISO 7886 

stethoscope icon

 

Hypodermic syringes are critical medical devices that allow medication to be delivered safely and efficiently at very precise doses. Because of their importance, it is crucial that hypodermic syringes perform as expected at all times. A leaking syringe can result in improper medication dosing and wastage and cause procedures to be delayed. The force required to expel liquid from the syringe must also be small enough to ensure that any medical professional can perform the task without resistance. 

ISO 7886-1 ANNEXES AND MEASUREMENTS


The ISO 7886-1 standard regulates the mechanical properties of hypodermic syringes and outlines the specifications for single use hypodermic syringes. It also includes information related to the design, manufacture, and functionality of the devices, while the annexes of the standard describe the standard operating procedure for testing these devices. Whereas ISO 7886-1:1993 discusses mechanical testing in annex G, the newer version of the standard, ISO 7886-1:2017 discusses it in annexes D and E.

Annex D measures the quality of the plunger stopper seal by simultaneously applying a perpendicular force to the syringe barrel and maintaining a constant compressive force on the syringe piston. A visual check is performed to ensure that no liquid escapes past the plunger stopper seal. This annex is less commonly tested than Annex E, which measures the forces needed to operate the syringe plunger with water in the barrel. The force profile typically displays an initial peak force known as the break loose force, and an average force of the remaining plunger travel, known as the glide force. This test also requires the recording of maximum force during plunger travel excluding the break loose force.

Testing System


ISO 7886-1 testing is performed on a single column universal testing system such as those found in Instron's 3400 or 6800 Series. A 2870-003 syringe fixture is required and is designed for this method specifically. Because the forces involved are relatively small, a 100 N or 50 N load cell is preferred. A biotray may also be helpful since there is potential for spillage from the reservoir onto the test system frame.




ISO 7886-1 Test Setup
1) 6800 Series Universal Testing System
2) 2530 Series Load Cell
3) 2870-003 Syringe Test Fixture

 

ISO 7886 test setup
Recent Changes to the Standard 

ISO 7886-1:2017 introduced many changes which require altering the test procedure and the results. The previous version of the standard required the expulsion and aspiration of the syringe while the newest version only requires expulsion similar to a typical break loose and glide force test. The fixturing has also changed, with the old method requiring the tubing from the syringe to be affixed to the reservoir. The new version requires a specific gauge and length needle to be attached to the end of the tube and then placed in the reservoir, resulting more repeatable results. To review the requirements in full, read the standard. To learn more about drug delivery device and container testing, click here

Related Content

6800 시리즈 프리미어 시험기 브로셔

Instron 6800 시리즈 만능 시험기는 최고의 정확도와 신뢰성을 제공합니다. 6800 시리즈는 완전히 새로운 Smart-Close Air Kit 및 Collision Mitigation 기능이 있는 특허 출원 중인 Operator Protect 시스템 아키텍처를 기반으로 하며, 그 어느 때보다 간편하고 스마트하며 안전한 소재 시험을 가능하게 합니다.

3400 시리즈 - 경제적인 시험 솔루션

인장, 압축, 굽힘 및 기타 소재 특성 시험을 위한 Instron 3400 시리즈 만능 시험기.

Bluehill Universal 브로셔

Bluehill Universal 소프트웨어는 터치 상호작용과 직관적인 사용자 경험을 위해 처음부터 새로 구축되었습니다. 사전 로드된 시험 방법, 몇 초 만에 진행되는 QuickTest, 향상된 데이터 내보내기 및 서비스에 대한 직접 통신 링크를 제공하는 신규 기능인 Instron Connect와 같은 기능으로 더 간단하고 스마트한 시험을 확인해보십시오. Bluehill 2 및 Bluehill 3와 같은 이전 버전의 소프트웨어 사용자는 Bluehill의 최신 버전으로 쉽게 업그레이드할 수 있습니다.

2870-003 주사기 시험 치구 - ISO 7886-1용

Instron® 주사기 시험 치구는 멸균, 일회용 피하 주사기에 대한 ISO 7886-1:2017 부칙 E 및 ISO 7886-1:1993 부칙 G의 시험 요구사항을 충족하도록 설계되었습니다.