ASTM F2458 Wound Closure Strength of Tissue Adhesives and Sealants

Mechanical Testing of Bioadhesives

ASTM F2458

ASTM F2458 is a testing standard that specifically focuses on characterizing the wound closure strength of tissue adhesives or sealants. Adhesive wound closure is sometimes preferable to using staples or stitches, as it reduces patient discomfort is and is often more easy to apply. It is extremely important that these products perform as expected, as failure could cause internal or external bleeding and leave a patient vulnerable to infection at the wound site. In order for wounds to heal properly over time and prevent scarring, the adhesive must continue to hold the tissue together until healing occurs.

ASTM F2458 is primarily intended for comparative testing between different adhesive formulations. Identifying a repeatable and life-like substrate enables this standard to be used for quality control purposes of adhesive wound closure products. The test will provide the maximum force at the time of failure as well as the failure mode – typically labelled as adhesive (failure of the adhesive/substrate bond), cohesive (failure of the internal adhesive bond), or substrate failure (failure of the tissue). To read the standard in full, purchase ASTM F2458.

Materials Testing System

For ASTM F2458 testing we recommend a 6800 Series single column testing system with a 100 N load cell and pneumatic grips. Because tissues are typically fragile, they require surfalloy style faces for clamping on the specimen successfully. A flexible coupling is required to ensure the loading is applied through the center of the specimen. For closure devices used internally within the body, the testing should be performed at a temperature of 37 Celsius. A biobox is useful for testing at body temperature.

ASTM F2458 Test Setup
1) 5900 Series Universal Testing System
2) 2530-100 Load Cell
3) Pneumatic Side Action Grips
4) Instron BioBox

 

ASTM F2458 system setup
Related Content

6800 Series Premier Testing Systems Brochure

Instron 6800 Series Universal Testing Systems provide unparalleled accuracy and reliability. Built on a patent-pending Operator Protect system architecture with an all-new Smart-Close Air Kit and Collision Mitigation features, the 6800 Series makes materials testing simpler, smarter, and safer than ever before.

Bluehill Universal Brochure

Bluehill Universal Software is built from the ground-up for touch interaction and an intuitive user experience. Discover simpler and smarter testing with features such as pre-loaded test methods, QuickTest in seconds, enhanced data exporting: and Instron Connect – a new feature that provides a direct communication link to Service. Users of previous versions of software such as Bluehill 2 and Bluehill 3 can easily upgrade to the newest version of Bluehill.

BioBox for 37°C Medical Device Testing

Instron's BioBox provides a controlled air environment at body temperature (37°C) for convenient testing of medical devices and biomaterials.

Submersible Side Action Pneumatic Grips and Temperature-Controlled BioBath

Instron's temperature-controlled BioBath allows for accurate simulation of the environmental conditions required for biomedical testing.