Every result your lab reports depends on the force reading. If the load cell reads high or low, every tensile strength, yield point, and modulus calculated from it is wrong, and an auditor or customer may reject the data. Regular force calibration proves your system measures force within the limits your standards allow.
Our factory-trained Field Service Engineers calibrate your system on site, to ASTM E4 and ISO 7500-1, under ISO/IEC 17025 accreditation. We were one of the first materials testing companies to use strain gauge load cells, and we still design and build our own. Since 2020, we have performed force calibrations on more than 41,000 load cells.
Choose Your Calibration Level
Both levels meet ASTM E4 and ISO 7500-1 in full, and both come with an accredited certificate. The difference is how low in the load cell’s range the calibration reaches. A load cell can only be used for results down to the lowest force it was calibrated at. Below that point, readings are outside the verified range, and an auditor can reject them.
A calibration that reaches lower forces lets one load cell cover more of your tests. You change load cells less often, which saves setup time and avoids the risk of mounting the wrong one. You may also not need to buy a second, smaller load cell for low-force tests.
Standard calibration suits labs that test one type of material in a narrow force range, such as a quality lab running the same tensile test every day. Enhanced calibration suits labs that test a wide range of materials on one system, for example metals at high forces and films, foams, or fibers at low forces.
Accredited to ISO/IEC 17025
NVLAP, the U.S. National Voluntary Laboratory Accreditation Program, has assessed our calibration methods, equipment, and staff competence against ISO/IEC 17025. Every force calibration within our scope comes with an accredited certificate your auditors will accept. View our scope of accreditation or see Calibration Lab and Accreditations.
Example: A 50 kN (11,240 lbf) load cell with standard calibration is verified down to 500 N (112 lbf) to ASTM E4. With enhanced calibration on a suitable system, it can be verified down to 50 N (11.2 lbf), so the same load cell can also cover low-force tests. For the other calibration services we offer, see Types of Calibrations.
What Happens During a Calibration
-
Schedule the VisitTell us your system, load cells, and the standard you need. We schedule a visit around your test plan to keep downtime short.
-
Record As-Found ReadingsThe engineer applies known forces with calibrated reference devices and records how your system reads them before any adjustment. This shows whether past results were within tolerance.
-
Adjust if NeededIf a reading is outside the limit, the engineer adjusts the system and repeats the check.
-
Verify As-Left ReadingsFinal readings across the calibrated range confirm the system meets ASTM E4 or ISO 7500-1. Our Calpro CR software records every reading automatically, so there are no manual transcription errors.
-
Receive Your CertificateYou receive an accredited calibration certificate for your auditors, also available in the Instron Connect portal with your service history.